Specifications


Number of EMG/EP channels 2, 4
Range of relative error of tension measuring on EMG channels in the range of input signals from 20 to 60000 mcV, no more than ±10%
Input impedance, no less than 100 MOm
Internal noise tension brought to input, no more than 5 µV
Amplitude of square calibration signal 5 mV ±5%
Range of relative error of time interval measuring , no more than ±2%
Overshoot of frequencyIn the range of frequency from 0,2 Hz to 3 kHz оfrom -30 to +10%
Time constant, no less than 1 s
Channel-to-channel attenuation coefficient, no less than 60 dB
CMRR on frequency 50Hz, no less than 100 dB
Sampling rate 20 000 Hz
Electrical stimulator:
-
current intensity
-
duration of stimulus

0.1...100 mA
0,06...2 ms
Operating system Windows 7, Vista, XP, 2000
PC connection USB
Electrical safety Class II type BF



 